Access keys

West Virginia Office of Emergency Medical Services

Medical Device Recall: LifePak CR Plus AED

Posted: Tuesday, December 09, 2008 11:07 AM

http://service.govdelivery.com/service/view.html?code=USDHSFA_11

September 11, 2008

Medical Device Recall: LifePak CR Plus Automated External Defibrillators

Audience: Emergency medical personnel, other healthcare professionals Physio Control, Inc., issued a recall of LifePak CR Plus Automated External Defibrillators (AED), used by emergency or medical personnel to treat adults in cardiopulmonary arrest. The product was recalled because the AED instructs the responder by voice prompts to press the shock button which is not visible because it is covered, thereby making the
responder unable to provide shock therapy. The AED device should be removed from service, or the manufacturer-provided diagram should be consulted to remove and discard the shock button cover.